Click to edit Master title style,*,Click to edit Master text styles,Second level,Third level,Fourth level,Fifth level,Internal use only,飞利浦医疗保健,zhang hua,Jun.29,2010,Click to edit Master title style,Click to edit Master text styles,Second level,Third level,Fourth level,Fifth level,*,Internal use only,飞利浦医疗保健,zhang hua,Jun.29,2010,Click to edit Master title style,Click to edit Master text styles,Second level,Third level,Fourth level,Fifth level,*,Internal use only,飞利浦医疗保健,zhang hua,Jun.29,2010,Click to edit Master title style,Click to edit Master text styles,Second level,Third level,Fourth level,Fifth level,*,飞利浦医疗保健,超声设备的维修和保养,2,内容,电网和用电平安,地线,日常维护,IU22 和IE33 飞利浦超声的简述,故障代码的分析和排除,3,正确使用交流电,提议:超声不要跟电脑、电风扇等其他设备共用一个插线板。,零线 火线,反 接,使用廉价插线板,能插就插,插满为止,不妥的行为和习惯:,无视接地,4,正确的接线,(,单项,220V,50Hz),左零、右火、中间地。,地线,黄绿花,火线,棕色,零线,蓝色,5,每个季度检查一次地线,请求助电工或飞利浦公司。,万用表是简易可行的方法,交流档检查零地电压,要求小于,2 V,。,ErrorId:0,File:D:Viewsusdbtlbca1ms5b5_g2b_rel_fec_svborisRoot borisHardwarefecappmainfecMonitor.cpp,Line number:1713,Info Msg:Error in calibrateHVi().Quiescent Current HVI-=0 x1ae=26 mA,FEC ASSERT(Print Only)-Expression:0,Info Msg:Unable to get semaphore in CFecOutMsg(),*Power Mon Calibration Failed!*,提示前端有过流的问题,但无法确认是哪块PCB,现场准备 最小配置隔离 AFP,NAIM、FEC,可以解决问题,34,Q&A,