Click to edit Master title style,Click to edit Master text styles,Second level,Third level,Fourth level,Fifth level,29,/,G,E,/,2024/9/16,MAC 800,心电分析系统,DCAR,系列,培训,闵昊,28,8,2009,课程导航:,MAC 800,安装指导,MAC 800,维修常见故障诊断与排除,MAC 800,安装指导,外观检查,检查外包装箱是否有碰撞、破损的痕迹。打开包装,检查机器,看是否外观有损坏。如果发现,请立刻通知,GE,公司的销售代表,并拍照取证;,对照装机清单,查看应有的配件是否齐全,电源线,患者缆线(导联线), 导联夹,吸球,锂电池,,SD,卡。如果有选配功能,查看是否相关配件齐全;,外观检查(前面板),外观检查(侧视),外观检查(后视),外观检查(底面),电池,键功能检查,装机流程,Insert the battery,Connect the AC power adapter,Connect,leadwires,Insert paper,Connect the LAN,Connect the external laser print,Turn on the unit,System Configurations,Test the equipment,Testing Entering Patient Information,Testing Resting ECG,Testing Arrhythmia Mode,Testing RR Analysis,Testing Intern Storage Management,Testing System Configuration,MAC 800,维修常见故障诊断与排除,故障一:无法开机,无屏显,Acquisitions,$1bn$2bn,GE capabilities,$0.5bn,Market growth,$0.5bn,交流供电时,检查220,V,交流电是否有输出;,电池供电时,检查是否电池电已耗尽,或损坏(更,换电池对比测试);,电源板(,AC-DC),,主板,液晶屏,键盘板,保险,丝均有可能损坏;,故障二:无法打印,或打一段卡纸,导联线损坏,或因操作导致的导联线与机器接触不良;,打印仓中折叠纸的放置方式错误,或者选用了尺寸不,合的打印纸;,打印仓门没有关好;,打印模块损坏;,打印模块的传动齿轮磨损;,主板损坏;,故障三:波形干扰较严重(基线漂移),导联接触不好,用酒精帮患者清洁皮肤并擦洗电极;,在设置菜单中开启,ADS,抗基线漂移功能;,开启肌电滤波功能;,调整滤波范围;,不和别的医疗产品共享同一个插座;,最好插在良好接地的插座上;,故障四:医生觉得打印波形异常,客户混淆了,IEC,和,AHA,导联标准,导致,接错了线;,导联线损坏;,主板损坏;,故障五:无法将数据传至,SD,卡,SD,卡处于,LOCK,状态,无法进行读写操作;,SD,卡损坏;,使用了容量超过2,GB,的,SD,卡;,SD,为格式化为,FAT,或,FAT16,格式;,SD,卡没有牢牢插入,SD,卡槽;,故障六:数据传到,PC,上却打不开,选择了,PDF,格式传输到电脑上,而电脑上去没有安,装,Adobe Reader,之类打开,PDF,文件的相应软件;,选择了,Hilltop,格式传输到电脑上,而电脑上没有安,装,GE,公司的,Cadiosoft,软件;,故障七:不能将数据导出到共享目录,确认是否已经激活,LANC,选项(选项设置),确认网络电缆已连接,确认通讯设置中,,IP,地址,子网掩码,网关,,DNS,服务器地,址均无误;,确认通讯设置中登陆信息无误(用户名,密码);,确认共享文件夹被设定为,Full control(,完全控制);,Thanks,